Supplies
Lancet | Drawer under exam counter (Blue and White) |
Alcohol wipe | Jar on room counter |
cotton ball | Jar on room counter |
Affinion Cartridge | Basket on wall in the lab behind lab door |
Add. Cartridges | Lab Refrigerator Door (must be room temperature to use) |
testing procedure
TESTING PROCEDURE
3.0 | Testing Procedure
Video Below
Collecting Sample:
Go to lab and
Turn on A1c machine if necessary
Self-Test will take up to 5 minutes
Power button on top of machine – towards back of device
Or wake up the far right machine
- Click on red top button
- Click red patient icon
- Enter patient’s chart number
- Click return button
- Patient’s chart number should pop up in the top left corner of the screen
Get A1c cartridge to collect blood sample
- In a black basket in the lab, in a plastic wrap.
- Extras are in the fridge – must be room temp before using them
- Keep 3 in basket at all times
- Let OM know it is time to order more when the last box is opened
- Remind OM when there are only 6 remaining in the fridge
- make sure they are standing upright
- PUT ON GLOVES
- Open the corner cap just slightly
- Ask the patient which finger they prefer to use
- Wipe off tip of finger with an alcohol wipe
- Allow finger to dry about 5 seconds
- Patient can shake finger to dry faster
- Twist and pull the safety tab on the lancet
- Place lancet with needle end against the side of the finger
- Do not place need in center of finger due to sensitive in that area
- Squeeze finger to get blood started
- Wipe first drop away with cotton ball
- Squeeze finger to get enough blood
- Use clear cartridge corner piece with small capillary tube to take up a small amount of blood
- Put back in a1c cartridge
- Give the patient a cotton ball to hold against finger prick
- Take to the lab immediately
- Insert cartridge into machine
- Barcode facing left
- Close the lid on A1c machine
- Machine will take 3-5 minutes to result
- Make note of result number to enter into chart
- Click checkmark
- Remove Cartridge
- Put cartridge and lancet in biowaste container
- Turn off machine at the end of day
Alere Afinion HbA1c Manual
It is a long established fact that a reader will be distracted by the readable content of a page when looking at its layout. The point of using Lorem Ipsum is that it has a more-or-less normal distribution of letters, as opposed to using ‘Content here, content here’, making it look like readable English. Many desktop
Afinion AS 100 analyzátor video
Afinion is a new concept in Point of Care – a multiparameter analyzer system. Afinion is designed to enable on-the-spot testing, using whole blood, plasma or urine samples. Physicians’ offices and clinics will find that Afinion saves time and increases productivity.
We help you design the life you want.
I am text block. Click edit button to change this text. Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo luctus nec ullamcorper mattis